Sampling Systems

Sample Recovery System (I1)

1. Off2. Recovery3. Vent4. Off

Provide a new septum on your sample bottle. Insert the bottle, with cap and septum into the slee­ve. Put the special bottle retaining clip in place and turn the thumb screw until the bottle is fixed in the sleeve.

Turn the sleeve upside down and fix the special bottle retaining clip in the terry clips. Turn the handle to the “recovery” position, allowing the inert gas to drive the liquid out of the sample bottle to the process-out connection. This position can be held for any required time.

Turn the handle to the "vent" position allowing the sample bottle to depressurise.

Turn the handle to the "off" position and turn the sleeve back to the normal position. Unfix the thumb screw, remove the bottle retaining clip and pull the bottle out from the sleeve. The septum reseals automa­tically.