S23 No Purge Configuration (D7)

The ability to purge the sample point ensures optimal retrieval of a representative liquid sample. The S23 type sampler in no purge configuration provides a system purge in addition to a fixed sample volume. No vent connection is needed. The vent is connected to the top valve. This way a loop is created in order to fill the bottle by means of gravity. This liquid sampler offers one handle operation by multiple valves, allowing for sampling independent of the process conditions. These features provide sample accuracy, safety for the operator and zero emission.

1. Off2. System Purge3. Sample4. Off

Provide a new septum on your sample bottle. Insert the bottle, with cap and septum into the sleeve until the septum is pierced by the needles. Put the bottle retaining clip in place.

Turn the handle to the "bypass" position, allowing a continuous flow of liquid through the sample chamber to ensure representative sampling. This position can be held for any required time.

Turn the handle to the "sample" position allowing the product to flow into the sample bottle. The fixed volume chamber determines the amount of sample. This position can be held for any required time.

Turn the handle to the "off" position. Remove the bottle retaining clip and pull the bottle out from the sleeve. The septum reseals automatically.